An Entirely Modular Detector Concept
- can be manufactured on industrial scales
- large detector area feasible (> 1 m2) at < 10% blind area
- time resolution < 1 ms (for TOF)
- based on GEM technology (invented by CERN)
Modern,
Highly Integrated ASIC Readout Electronics
- individual readout channels
- simple to scale
- parallel processing of highest count rates
Suited For Highest Intensity Applications
|
|
CASCADE Detector,
200 mm x 200 mm,
stack of GEM-foils with 2D-readout-structure in the middle,
complete readout and histogramming electronics at the backside of the detector
|
|
Signal Rates Capacity
- Potential of GEMs:
up to 10 MHz/cm2
- Experimentally verified to date:
> 2.7 MHz/cm2
> 4 MHz on detector
Detection Efficiency
- 38% at 1.8 Angström predicted for 20 10Boron layers coated on GEM-Foils
- Measured 23% at 2.5 Angström using 8 layers of 10Bor
- Custom optimisation for predefined band of wavelength possible
- Black detector beyond 5 Angström
Spatial Resolution
- feasible resolution:
1-5 mm depending upon counting gas, readout structure and detector design
- Measured width: FWHM 2.6mm on 1.57mm stripes using Ar/CO2 = 90/10 as counting gas at
normal pressure
|